Sample report check
What a check looks like when it comes back: each mismatch with the page it is on, what the other documents say and what to look at. Your reviewer decides what to change.
- Job
- Example Station, Engine 2 outlet (invented)
- Unit
- Natural gas engine, 1,380 hp
- Methods
- EPA Methods 3A, 7E and 10; 3 runs
- Checked
- Final report (96 pages), field data workbook, ERT project file
- Result
- 3 mismatches, 1 question, 1 wording note. Everything else agrees.
Mismatches
-
Mismatch
Run 2 NOx at 15% O2: the ERT file differs from the report
Source Where Value, ppmvd Report, results table p. 14 7.96 Field data workbook, DAS export row 418 7.96 ERT file, run data Run 2, NOx 9.76 The report and the field data agree. The ERT entry looks like the same digits in another order. The test average in the ERT file changes with it.
-
Mismatch
Average CO does not follow from the three runs
Run Where CO, ppmvd at 15% O2 Run 1 p. 22 41.2 Run 2 p. 22 42.9 Run 3 p. 22 42.8 Average of the runs, recomputed 42.3 Average as printed p. 22 41.8 The printed average is 0.5 ppm lower than the average of the printed runs. The ERT file carries the runs, so ERT will show 42.3.
-
Mismatch
NOx mid-level gas: the certificate and the calibration sheet disagree
Source Where Certified value, ppm Cylinder certificate, cylinder EX-0412 (invented) p. 61 45.2 Calibration sheet p. 58 54.2 ERT file, gas cylinder list EX-0412 54.2 The certified value enters the calibration error, the system bias and every bias-corrected run result. With the certificate's 45.2 ppm, the pre-test calibration error for this gas is outside the Method 7E limit of 2.0 percent of span.
Question
-
Question
Was the engine tested at the load the rule asks for?
The report gives the engine load as 82 percent during all three runs (p. 9). 40 CFR 63.6620(b) asks for a test within plus or minus 10 percent of 100 percent load for this engine class. The report does not say why the load was lower. There may be an approval we can't see.
Wording
-
Wording
"Turbine" in an engine report
The introduction on p. 3 says "the turbine was operated at normal load". Every other page describes a reciprocating engine. This reads like text carried over from another job.
Checked, nothing to report
- Runs 1 and 3: report, field data and ERT file agree for NOx, CO and O2
- Calibration error, system bias and drift for O2 and CO, against the Method 7E limits
- Gas cylinders certified before the test and not expired
- Every calibration in the ERT file uses a cylinder on the cylinder list
- Run start and end times in order on the test day
- Units fit each gas: O2 in percent, NOx and CO in ppm
- O2-corrected values follow from the measured ones
- Every field ERT requires is filled in
Stackbook never declares compliance and never signs or submits. Your reviewer decides on each item and your firm authorizes the report.